Changes of endothelial function and intracardiac hemodynamics during implantation of different types of pacemakers

https://doi.org/10.15829/1560-4071-2019-8-36-43

Abstract

Aim. To determine the dynamics of endothelial function and intracardiac hemodynamics during implantation of different types of pacemakers.

Material and methods. The study involved 61 patients, 29 of them were implanted with single-chamber pacemaker (the first group), 32  — two-chamber pacemaker (second group) with sick sinus syndrome (SSS), as well as patients with 2-3 degrees AV conduction disorder. Test with endothelium-dependent vasodilation of the brachial artery was performed using the standard technique and the velocity parameters of blood flow were determined. Parameters of intracardiac hemodynamics and left atrial (LA) function were determined. These studies were conducted in patients initially upon admission to the hospital and 2 months later implantation of pacemaker.

Results. Endothelium-dependent vasodilation significantly deteriorated in patients with single-chamber pacemaker which initially was 6,1 [2,74;9,815]% and in 2 months decreased to 5,4 [2,37;7,55]% (p=0,05). There was a tendency to deterioration of left ventricular (LV) contractile function accompanied by a decrease in ejection fraction (EF) to 55,04 [49,04;68,72]%, and decrease emptying fraction of LA to 37,36 [31,67;52,85]%. There was a tendency to increase the volume index of LA which indicates an increasing overload of the volume and pressure of LA. There was a significant increase in the degree of endothelium-dependent vasodilation in patients of the second group initially compiling a 3,25 [2,15;6,61]% and then reaches 6,15 [2,75;8,97]% 2 months after surgery (p<0,04). There was no overload in the volume and pressure of LA and improved indirect signs of LV compliance parameters (growth of end-systolic dimension and volume) and decreased LV EF to 53,23 [44,1;69,95]%. In patients in the first group there was a tendency to increase peripheral resistance to 1227 [1065;1318] dyne x cm-5 x sec, accompanied by a significant decrease in blood flow rate. In the second group there was a tendency to reduce peripheral resistance to 1215 [1136;1391] dyne x cm-5 x sec and a significant increase in blood flow rate. There was a significant correlation between the systole LA dimension and total peripheral resistance in patients with VVI stimulation (r=0,78, p=0,0022), which indicates an increase in total peripheral resistance with a change in blood flow from LA to LV and a decrease in cardiac output.><0,04). There was no overload in the volume and pressure of LA and improved indirect signs of LV compliance parameters (growth of end-systolic dimension and volume) and decreased LV EF to 53,23 [44,1;69,95]%. In patients in the first group there was a tendency to increase peripheral resistance to 1227 [1065;1318] dyne x cm-5 x sec, accompanied by a significant decrease in blood flow rate. In the second group there was a tendency to reduce peripheral resistance to 1215 [1136;1391] dyne x cm-5 x sec and a significant increase in blood flow rate. There was a significant correlation between the systole LA dimension and total peripheral resistance in patients with VVI stimulation (r=0,78, p=0,0022), which indicates an increase in total peripheral resistance with a change in blood flow from LA to LV and a decrease in cardiac output.

Conclusion. Endothelial function deteriorates with single-chamber stimulation and improves with two-chamber stimulation. These changes are based on shear stress and nitric oxide endothelium secretion the changes of which depend on the characteristics of central and peripheral hemodynamics.

About the Authors

V. A. Sanakoeva
Russian Medical Academy of Continuing Professional Education
Russian Federation
Moscow
Competing Interests:

Конфликт интересов не заявляется.



R. Z. Pashaev
Consultation and Diagnostic Center of the V.V. Veresaev City Clinical Hospital
Russian Federation
Moscow
Competing Interests: Конфликт интересов не заявляется.


A. A. Pukhaeva
Russian Medical Academy of Continuing Professional Education
Russian Federation
Moscow
Competing Interests: Конфликт интересов не заявляется.


A. G. Avtandilov
Russian Medical Academy of Continuing Professional Education
Russian Federation
Moscow
Competing Interests: Конфликт интересов не заявляется.


References

1. Shalnova SA, Konradi AO, Karpov YA, et al. Сardiovascular mortality in 12 russian federation regions — participants of the “cardiovascular disease epidemiology in Russian regions” study. Russian Journal of Cardiology. 2012;(5):6-11. (In Russ.)

2. Bokeriya LA, Revishvili ASh, Dubrovskij IA. State of electrocardiostimulation in Russia in 2010. Vestnik aritmologii. 2012;(68):77-80. (In Russ.)

3. Abdul’yanov IV, Vagizov II. Modern approaches to permanent cardiac pacing. Prakticheskaya medicina. 2013;71(3):49-54. (In Russ.)

4. Krishnamoorthy S, Khoo CW, Lim HS, et al. Predictive value of atrial high-rate episodes for arterial stiffness and endothelial dysfunction in dual chamber pacemaker patients. Eur J Clin Invest. 2014;44(1):13-21. doi:10.1111/eci.12182.

5. Revishvili ASh, Boytsov SA, Davtyan KV, et al. Clinical guidelines for electrophysiological studies, catheter ablation and the use of implantable antiarrhythmic devices. Novaya redakciya, 2017. 702 p. (In Russ.)

6. Matusik PT. Adverse Clinical Outcomes Related to Right Ventricular Pacing. European Heart Journal. 2019;40(20):1586-8. doi:10.1093/eurheartj/ehz279.

7. Cho SW, Gwag HB, Hwang JK, et al. Clinical features, predictors, and long-term prognosis of pacing-induced cardiomyopathy. Eur J Heart Fail. 2019;21(5):643-51. doi:10.1002/ejhf.1427.

8. Kiehl EL, Makki T, Kumar R, et al. Incidence and predictors of right ventricular pacinginduced cardiomyopathy in patients with complete atrioventricular block and preserved left ventricular systolic function. Heart Rhythm. 2016;13(12):2272-8. doi:10.1016/j. hrthm.2016.09.027.

9. Khurshid S, Epstein AE, Verdino RJ, et al. Incidence and predictors of right ventricular pacing-induced cardiomyopathy. Heart Rhythm. 2014;11(9):1619-25. doi:10.1016/j. hrthm.2014.05.040.

10. Dong YX, Guo M, Yang YZ, et al. Effects of ventricular demand and dual-chamber pacing models on the long-term clinical outcome and cardiac remodeling in patients with symptomatic bradycardia. National Medical Journal of China. 2011;91(30):2103-7. doi:10.3760/cma.j.issn.0376-2491.2011.30.005.

11. Zhang X, Li Y, Wang N, et al. Effects of permanent cardiac pacemaker implantation on vascular endothelial function, blood coagulation and cardiac function in patients with bradycardia. Exp Ther Med. 2018;16(6):4717-21. doi:10.3892/etm.2018.6808.

12. Domenichini G, Sunthorn H, Fleury E, et al. Pacing of the interventricular septum versus the right ventricular apex: A prospective, randomized study. Eur J Intern Med. 2012;23(7):621-7. doi:10.1016/j.ejim.2012.03.012.

13. Kypta A, Steinwender C, Kammler J, et al. Long-term outcomes in patients with atrioventricular block undergoing septal ventricular lead implantation compared with standard apical pacing. Europace. 2008;10(5):574-9. doi:10.1093/europace/eun085.

14. Kostogryz VB, Turovskaya TV. Features of nitric oxide metabolism in patients with chronic heart failure of ischemic etiology with implanted pacemaker: possibilities of drug correction by beta-adrenoceptor blockers. Ukr. med. CHasopis. 2010;5(79):81-4. (In Russ.)

15. Rubaj A, Ruciński P, Rejdak K, et al. Biventricular versus right ventricular pacing decreases immune activation and augments nitric oxide production in patients with chronic heart failure. Eur J Heart Fail. 2006;8(6):615-20. doi:10.1016/j.ejheart.2005.12.001.

16. Choy AM, Su HH, Elder DH, et al. Right ventricular pacing impairs endothelial function in man. Europace. 2011;13(6):853-8. doi:10.1093/europace/eur028.

17. Celermajer DS, Sorensen KE, Gooch VM, et al. Non-invasive detection of endothelial dysfunction in children and adults at risk of atherosclerosis. Lancet. 1992;340(8828):11115. doi:10.1016/0140-6736(92)93147-f.

18. Golicyn SP. Changes in hemodynamic parameters and neurohumoral regulation systems in patients with implanted pacemakers after switching the DDD stimulation mode to VVI mode. Cardiology. 2006;(11):66-70. (In Russ.)


Review

For citations:


Sanakoeva V.A., Pashaev R.Z., Pukhaeva A.A., Avtandilov A.G. Changes of endothelial function and intracardiac hemodynamics during implantation of different types of pacemakers. Russian Journal of Cardiology. 2019;(8):36-43. (In Russ.) https://doi.org/10.15829/1560-4071-2019-8-36-43

Views: 676


Creative Commons License
This work is licensed under a Creative Commons Attribution 4.0 License.


ISSN 1560-4071 (Print)
ISSN 2618-7620 (Online)